Breath Well Medical Systems
Breath Well Medical Systems
Konanakunte, Bengaluru, Karnataka
GST No. 29ABOPU7704Q1ZH

Resmed Ventilators

We are a leading Wholesale Trader of resmed lumis 100 vpap s non-invasive ventilator, resmed lumis 150 vpap st non-invasive ventilator, resmed stellar 150 non-invasive ventilator and resmed stellar 100 non-invasive ventilator from Bengaluru, India.
Request CallBack
Resmed Lumis 100 VPAP S Non-Invasive Ventilator
Get Best Quote

Resmed Lumis 100 VPAP S Non-Invasive Ventilator

Approx. Price: Rs 80,300 / NumberGet Latest Price
Product Details:
BrandResmed
ModesS
Usage/ApplicationHospital
Operating Mode28010
Specifications
Height150 mm
Weight1106 gm
Breadth205 mm
Length116 mm
ModesCPAP, S
Operating pressure range2–25 cm H2O (2–25 hPa) in S mode 4–20 cm H2O (4–20 hPa) in CPAP mode
Sound pressure level26.6 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2009 (CPAP mode)
Dimensions (H x W x D)116 mm x 255 mm x 150 mm (Device with HumidAir humidifier)
Weight (Device with HumidAir humidifier)1268 gm
Power Supply90W power supply unit Input range: 100–240V, 50–60Hz, 57VA (typical power consumption) 108VA (peak power consumption)
Operating temperature/ humidity/altitude+5°C to +35°C/ 10–95% relative humidity, non-condensing/ Sea level to 2,591 m; air pressure range 1013 hPa to 738 hPa
Storage and transport temperature/humidity-20°C to +60°C/ 5–95% relative humidity, non-condensing
Housing constructionFlame retardant engineering thermoplastic
Supplemental oxygenRecommended maximum flow: 15 L/min (CPAP, S)
Standard air filterPolyester non-woven fibre
TubingSlimLine™ and ClimateLineAir (15 mm); Standard (19 mm)
Air outlet22 mm air outlet complies with ISO 5356-1:2004
Electromagnetic compatibilityRequirements (EMC) according to IEC60601-1-2:2007 for residential, commercial, and light industry environments
Aircraft useResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air travel
IEC 60601-1: 2006 classificationClass II (double insulation), Type BF Ingress protection IP22

Yes! I am interested
View PDF
Request CallBack
Resmed Lumis 150 VPAP ST Non-Invasive Ventilator
Get Best Quote

Resmed Lumis 150 VPAP ST Non-Invasive Ventilator

Approx. Price: Rs 1.1 Lakh / NumberGet Latest Price
Product Details:
BrandResmed
ModesiVAPS
Usage/ApplicationHospital
Operating Mode28123

 

Key Features - 
  • iVAPS
  • intelligent Backup Rate (iBR)
  • AutoEPAP maintains upper airway patency
  • Vsync
  • TiControl
  • 5 Trigger and Cycle Senstivities
  • QuickNav for low touch therapy adjustments
  • Climate Control Auto for automatic humidification
  • Ramp and Rampdown for extra comfort.
  • supported by ResScan 5.4

 

Specifications
Height150 mm
Weight1268 gm
Breadth255 mm
Length116 mm
Sound pressure level27 dBA with uncertainty of 2 dBA
Standard air filterMaterial: Polyester non woven fibre Average arrestance: >75% for ~7 micron dust
Aircraft useResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air travel.
ModesCPAP, S, T, ST with optional iBR, PAC, iVAPS with iBR and optional Auto EPAP
Water capacityTo maximum fill line 380 mL
SensorsPressure sensor: Internally located at device outlet, analogue gauge pressure type, -5 to +45 cm H2O (-5 to +45 hPa) Flow sensor: Internally located at device inlet, digital mass flow type, -70 to +180 L/min
Maximum single fault steady pressureDevice will shut down in the presence of a single fault if the steady state pressure exceeds: 30 cm H2O (30 hPa) for more than 6 sec or 40 cm H2O (40 hPa) for more than 1 sec.
Housing constructionFlame retardant engineering thermoplastic
Cleanable humidifiermaterial: Injection moulded plastic, stainless steel and silicone seal
Operating pressure rangeS, ST, T, PAC, iVAPS: 2 to 25 cm H2O (2 to 25 hPa) CPAP 4 to 20 cm H2O (4 to 20 hPa)
Operating temperature/ humidity/altitude+5°C to +35°C / 10 to 95% relative humidity, non-condensing / Sea level to 8,500' (2,591 m); air pressure range 1013 hPa to 738 hPa
Storage and transport temperature/humidity-20°C to +60°C / 5 to 95% relative humidity, non-condensing
Supplemental oxygenMaximum flow: 15 L/min (S, ST, T, PAC, CPAP), 4 L/min (iVAPS)
Air outlet(complies with ISO 5356-1:2004): 22 mm
Electromagnetic compatibilityrequirements (EMC) according to IEC 60601-1-2:2014, for residential, commercial and light industry environments. It is recommended that mobile communication devices are kept at least 1 m away from the device.
IEC 60601-1:2005+A1:2012 classificationClass II (double insulation), Type BF, Ingress protection IP22

Yes! I am interested
View PDF
Request CallBack
Resmed Stellar 150 Non-Invasive Ventilator
Ask For Price

Resmed Stellar 150 Non-Invasive Ventilator

Product Price : Get Latest Price
Product Details:
BrandResmed
ModesS/T
Usage/ApplicationHospital
Operating Mode24162
Height120 mm
Weight2.1 kg
Breadth170 mm
Length230 mm
Sound pressure level31 dBA with uncertainty of 3 dBA as measured according to ISO 17510 – 1:2007.
Maximum single fault steady pressure60 cm H2O (in all modes)
Breathing resistance under single fault2 cm H2O at 30 L/min; 7.2 cm H2O at 60 L/min
Electromagnetic compatibilityProduct complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial, and light industry environments.
Aircraft useMedical-Portable Electronic Devices (M-PED) that meet the Federal Aviation Administration (FAA) requirements of RTCA/DO-160 can be used during all phases of air travel without further testing or approval by the airline operator. ResMed confirms that the Stellar meets RTCA/DO-160 requirements
Storage and transport humidity10%–95% non-condensing
TubingFlexible plastic, 6’ 6” (2 m) or 9’ 10” (3 m) length (0.9”/22 mm diameter) SlimLine air tubing Flexible plastic, 6’ (1.83 m) length (0.6”/15 mm diameter
Air outlet0.9” (22 mm) taper, compatible with ISO 5356-1:2004 Anaesthetic & Respiratory Equipment - Conical Connectors
External DC Power Supply24 V, 3 A
Screen ResolutionDimension (L x W x H): 3” x 2.5” x 0.1” (76.9 mm x 63.9 mm x 3.15 mm) Resolution: 320 x 240 pixels
Operating pressure range•IPAP: 3 cm H2O to 40 cm H2O (in S, ST, T, PAC mode) •PS: 0 cm H2O to 37 cm H2O (in S, ST, T, PAC mode) •EPAP: 3 cm H2O to 25 cm H2O (in S, ST, T, iVAPS, PAC mode) •CPAP: 4 cm H2O to 20 cm H2O (in CPAP mode only) •Min PS: 0 cm H2O to 20 cm H2O (in iVAPS mode) •Max PS: 0 cm H2O to 30 cm H2O; 8 cm H2O to 30 cm H2O when AutoEPAP is on (in iVAPS mode) •Min EPAP/Max EPAP: 3 cm H2O to 25 cm H2O (in iVAPS mode when AutoEPAP is set to on) Note: iVAPS mode is only available in Stellar 150.
IEC 60601-1Class II (Clause 3.14—double insulation). This adherence means the need for an protective earthing (ie, an earthed plug) is not necessary. •Type BF •Continuous operation
Operating humidity10%–95% non-condensing
Pressure measurementInternally mounted flow transducer
Standard air filterElectro static fibre mesh with TPE frame structure. Bacterial filtration efficiency of 99.540% on area weight 100g/m².
Flow measurementInternally mounted pressure transducer
ModesCPAP, S, T, ST with optional iBR, PAC, iVAPS with iBR and optional Auto EPAP
Internal BatteryLithium-Ion battery, 14.4 V, 1.6 Ah, 23 Wh Operating hours: 2 h with a new battery under normal conditions (see below). Patient type: home chronic; pressure: IPAP/EPAP 15/5 cm H2O; mask type: Ultra Mirage; air tubing: 6’ 6” (2 m); leak: 0; respiratory rate: 20 bpm; battery capacity: 100% Patient type: hospital acute; pressure: IPAP/EPAP 20/5 cm H2O; mask type: Ultra Mirage; air tubing: 6’ 6” (2 m); leak: 0; respiratory rate: 45 bpm; battery capacity: 100%
Power supply unitAC 100–240V, 50–60Hz, 2.2 A, max. 65 W
Storage and transport temperature-4°F (-20°C) to +140°F (60°C); +122°F (+50°C*)

Yes! I am interested
View PDF
Save time! Get the Best Deal

*Your information is safe with us.

Request CallBack
Resmed Stellar 100 Non-Invasive Ventilator
Ask For Price

Resmed Stellar 100 Non-Invasive Ventilator

Product Price : Get Latest Price
Product Details:
BrandResmed
ModesS/T
Usage/ApplicationHospital

 

Key Features - 
  • Pathology defaults for obstructive, restrictive, obesity hypoventilation and normal lung mechanics
  • Invasive and non invasive, adult to peadiatric usage
  • In built battery back up of upto 2 hours
  • 2 customizable pre set programs
  • Vsync.
  • TiControl
  • 5 Trigger and Cycle Senstivities
  • Oxygen delivery upto 30 L/min
  • Comprehensive range of alarms
  • Real time monitoring and multiple graphs
  • Integrated SpO2 and FiO2 monitoring
  • Integrated Humidification

 

Specifications
Height120 mm
Weight2.1 kg
Breadth170 mm
Length230 mm
Sound pressure level< 30 dB(A)
Storage and transport temperature/humidity
Maximum single fault steady pressure60 cm H2O
Electromagnetic compatibilityProduct complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial, and light industry environments.
Aircraft useProduct complies with RTCA/DO-160 section 21 requirements. Consult the medical services department of your carrier if you intend to use your device on an aircraft. Note: You should not use the Stellar 100 while the aircraft is taking off or landing
Max Flow> 200 L/min at 20 cm H2O
TubingFlexible plastic, 6’ 6” (2 m) or 9’ 10” (3 m) length (0.9”/22 mm diameter) / SlimLine air tubing Flexible plastic, 6” (1.83 m) length (0.6”/15 mm diameter)
Air outlet0.9”(22 mm) taper, compatible with ISO 5356-1:2004 Anaesthetic & Respiratory Equipment - Conical Connectors
Operating pressure rangeIPAP: 3 cm H2O to 40 cm H2O • EPAP: 3 cm H2O to 25 cm H2O • CPAP: 4 cm H2O to 20 cm H2O
Standard air filterElectro static fibre mesh with TPE frame structure
Operating temperature/ humidity/altitude+32°F (0°C) to + 95°F (35°C) / 10%–95% non-condensing /
ModesCPAP, S, ST, T, PAC
IEC 60601-1Class II (double insulation), type BF, continuous operation
Internal BatteryLithium-Ion battery, 14.4 V, 1.6 Ah, 23 Wh Battery run-time: 2 hours (measured at IPAP15 cm H2O, EPAP 5 cm H2O, nasal mask, 6’ 6” (2 m) air tubing, no leak, 20 bpm)
Power supply unitAC 100–240V, 50–60Hz, 2.2 A, max. 65 W
Housing constructionFlame retardant engineering thermoplastic
Breathing resistance under single fault2 cm H2O at 30 L/min; 7.2 cm H2O at 60 L/min
Mask pressure toleranceIPAP: ± 0.5 cm H2O ± 10% of set pressure (end of inspiration) EPAP/PEEP: ± 0.5 cm H2O ± 4% of set pressure CPAP: ± 0.5 cm H2O ± 10% of set pressure
Mask flow tolerance± 6 L/min or 20%, whichever is greater
Pressure measurementInternally mounted pressure transducer
Flow measurementInternally mounted flow transducer
External DC Power Supply24 V, 3 A

Yes! I am interested
View PDF
X